Page 162 - 《高原气象》2026年第5期
P. 162
高 原 气 象 45 卷
1378
Temperature layout and testing in the double-75-m
3
expansion-type cloud chamber
Deng Yupeng, Liu Zhitao, Wang Li, Yang Yang, Li Zhenghao, Zhao Liwei
(Hebei Provence Weather Modification Center, Shijiazhuang 051430, Hebei, China)
Abstract: By setuping platinum resistance temperature sensors (PT100) and optical fibre temperature sensors
(OS), a distributed temperature observation system was constructed in an expansion-type cloud chamber, en‐
abling optimized spatial monitoring of the chamber temperature field. The spatio-temporal evolution of the gas
temperature distribution and the influence of wall temperature on air temperature in the enclosed space of the
cloud chamber was systematically investigated by conducting a series of associated experiments with two typical
conditions, thermally stable and dynamic expansion conditions. The results show that: (1) during the condition
of thermally stability, the temperature inside the chamber mainly changed through the heat transfer mode of the
inner wall of the chamber. Under static equilibrium conditions, the vertical gas temperature difference within 120
min was ≤0.1 ℃, and the horizontal difference was ≤0.05 ℃; during the uniformly temperature-controlling pro‐
cess from 0 to -40 ℃, the initial deviation between the upper and the lower temperatures in the chamber was
more than 4 ℃ due to the influence of the wall temperature, and then gradually reduced to <2 ℃;the rapid tem‐
perature-control process from 27 °C to -20 °C took approximately 78 min, with a cooling rate of 0.6 °C·min⁻¹,
exhibiting a fast-to-slow trend. (2) Under the condition of environmental change, i.e. dynamic gas expansion in
the cloud chamber, the maximum value of the horizontal gas temperature difference in the chamber was 2 ℃,
and the vertical gas distribution of the temperature presented a significant gradient characteristic, as temperature
systematically increasing with height within the chamber: the temperature was slightly higher in the upper part
and decreased towards the lower part. This conclusion reveals the dynamic response characteristics of gas temper‐
ature distribution under different thermodynamic conditions and its relationship with the gas expansion, provid‐
ing multidimensional empirical evidence for the precise characterisation of the thermodynamic parameters of the
experimental subject. (3) During the expansion process, the gas was heated by the chamber wall, which was no
longer controlled and maintained at a constant temperature, resulting in a real cooling rate significantly lower
than the theoretically calculated adiabatic expansion cooling rate. The time constant for the heating of the gas
within the chamber—that is, the relaxation time required for the gas to return to a steady state following dynamic
changes, is approximately 500 s. This parameter characterizes the reheating efficiency of the cloud chamber wall
to the internal gas after the expansion process, and directly affects the effect of cloud expansion and the mainte‐
nance time of cloud.
Key words: expansion-type cloud chamber; lab experiment; temperature layout; temperature testing; relaxing
time

